Calibration Service
Precise work with measurement technology requires technically flawless equipment. However, changing ambient conditions and thermal, mechanical or chemical stresses can lead to inaccurate measurements as the devices are used over a longer period of time.
In order to ensure the reliability of the measurement results and to ensure the longevity of the measuring devices, we recommend regular calibration. For this we offer you our calibration service with test certificate. The acceptance test certificate according to DIN EN 10204:2004/3.1 shows the data of the reference measurement, the calibration values and the measurement uncertainty, confirms the connection to national measurement variables and thus certifies the quality of your measuring equipment.
Calibration means checking the correctness of the measured variables of a measuring device or determining measurement deviations. The calibration takes place without any technical intervention in the measuring system.
Advantages
|
Calibration Items
Absolute Pressure | Barometers Barographs Air pressure sensors Air pressure data loggers |
Temperature | Thermometer Thermographs Temperature sensors Temperature data loggers |
Moisture | Hygrometer Hygrographs Humidity sensor Humidity data logger |
Temperature/Humidity | Thermal hygrometer Thermal hygrograph Temperature/humidity sensor Temperature/humidity data logger |
Test facilities
Test field for complex climate simulation (Fischer climate pressure test chamber)
Workspaces | Pressure | 20...1200 hPa |
Temperature | -40...+120 °C | |
Humidity | 5...98% RH (5...80% r.h. in the range -40...0 °C) | |
Change speed | Pressure | 100 hPa/min (100 ... 0.1 hPa/min in the range below 600 hPa) |
Temperature | 5 K/min (-40 to +120 °C), 2 K/min (+120 to -20 °C), | |
<2 k="" min="" -20="" after="" -40="" c="" td="" gt="" --2--=""> | ||
Humidity | 10%/min (5 after 98% RH), 3%/min (98 after 5% RH) | |
Stability of climate values | Pressure | ± 0.3 hPa |
Temperature | ± 0.1K | |
Humidity | ± 3% RH | |
reference gauges | Pressure | Pressure DPI 740 SNo. 74001562 Accuracy ± 0.15 hPa |
Temperature | DE P655 SNr 6550206 1060/020102/109014/050285 (± 0.03 K) | |
Humidity | DE P655 SNr 6550206 1060/020102/109014/050285 (± 1.5% RH) | |
Test space dimensions | 45 cm x 45 cm x 50 cm, approx. 100 l test chamber volume |